TECHNICAL PARAMETERS
Part of Anesthesia
* Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
* Scope of application | adult, child |
* Gas source | O2, N2O: 0.27~0.55MPa |
* Flow meter | O2: 0.1~1.0L/min 1.1~10L/min |
* Flow rate of rapid oxygen supply | 25~75L/min |
* Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
* Vaporizer |
It has the function of automatic compensation based on pressure,temperature,and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer |
* Respiratory circuit |
working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa |
* Respiratory bellows |
bellows for adults, bellows for children tidal volume range:0~1,500 ml |
Part of Ventilator
* Display mode | Digital display | |
* Ventilation mode |
IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate) |
|
* Ventilation parameters | Tidal volume | 50~1500ml |
Rate | 4~80 bpm | |
I:E | 2:1~1:4 | |
Pressure range | 1.0~6.0 kPa | |
* Parameters for ventilation monitoring |
tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform | |
Airway pressure alarm |
upper limit setting range 1.0~6.0 kPa low limit setting range 0.4~2.0 kPa |
|
TV alarm |
upper limit setting range 0.5~2.0L low limit setting range 0.02~1.0L |
|
Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa | |
Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds | |
Power alarm | ||
* Power | AC 220V±10%; 50Hz±1Hz; (UPS, Storage battery) |